Adjust The Echo Tracking Settings - Emerson 5301HxxxxxxxxxxxxxxZZ Series Reference Manual

Guided wave radar level transmitter
Table of Contents

Advertisement

Advanced configuration
April 2019
Reset DC estimation
Resets DC estimation to the configured initial value, forcing the device to start over
estimating the product DC.
Use static product DC
Check this setting if you do not want the device to estimate the product DC. This will force
the device to use the configured initial product DC.
D.4

Adjust the echo tracking settings

Measurement with the Rosemount 5300 is based on the fact that the radar signal pulses
are reflected by the product surface. Different parameters are used to track the
measurement signal to achieve a reliable and stable measurement. Normally, the echo
tracking parameters are automatically set by the transmitter, and no manual settings are
needed. However, due to the properties of the product, in rare cases it may be necessary
to adjust the echo tracking settings for optimum measurement performance.
Procedure
1. In Rosemount Radar Master, select Setup → Advanced.
2. Select the Echo Tracking tab.
3. To adjust the echo tracking settings, clear the Use automatic echo tracking
4. Set Echo Timeout and Close Distance to the desired values.
402
settings checkbox.
Figure D-4: Echo Tracking in Rosemount Radar Master
Reference Manual
00809-0100-4530
Rosemount 5300 Level Transmitter