Specifications; Educational Services - Emerson Fisher 4211 Instruction Manual

Electronic position transmitters
Hide thumbs Also See for Fisher 4211:
Table of Contents

Advertisement

Instruction Manual
D200354X012
Table 4. Fisher 4211 and 4221 Position Transmitter EMC Summary Results—Immunity
Port
Electrostatic Discharge (ESD)
Radiated EM field
Enclosure
Rated power frequency magnetic
field
Burst (fast transients)
Surge
I/O signal/control
Conducted RF
Specification limit = ±1% of span
1. A = No degradation during testing. B = Temporary degradation during testing, but is self‐recovering.
For instruments equipped with position switches, the high relay position switch can be set at any point of travel. The
low relay position switch can be offset from the high relay position switch trip point by as little as 1/20 of the travel. In
the event of a power loss both position switch relays de‐energize, indicating loss of transmitter power.

Specifications

Specifications for the 4200 transmitters are listed in table 1.
WARNING
This product is intended for a specific current range, temperature range and other application specifications. Applying
different current, temperature and other service conditions could result in malfunction of the product, property damage or
personal injury.

Educational Services

Emerson Automation Solutions
Educational Services - Registration
Phone: +1-800-338-8158
E-mail: education@emerson.com
emerson.com/mytraining
Phenomenon
Basic Standard
IEC 61000‐4‐2
IEC 61000‐4‐3
IEC 61000‐4‐8
IEC 61000‐4‐4
IEC 61000‐4‐5
IEC 61000‐4‐6
Test Level
4 kV contact
8 kV air
80 to 1000 MHz @ 10V/m with 1 kHz AM at 80%
1400 to 2000 MHz @ 3V/m with 1 kHz AM at 80%
2000 to 2700 MHz @ 1V/m with 1 kHz AM at 80%
60 A/m at 50 Hz
1 kV
1 kV (line to ground only, each)
150 kHz to 80 MHz at 3 Vrms
1 kHz AM at 80%
4200 Transmitters
September 2022
Performance
(1)
Criteria
A
A
A
A
B
A
5