International - Emerson 5301HxxxxxxxxxxxxxxZZ Series Reference Manual

Guided wave radar level transmitter
Table of Contents

Advertisement

Product certifications
April 2019
3. Parts of the sensor probes, for type 5300 are non-conducting material covering
4. The Ex ia version of model 5300 FISCO device may be supplied by an "Ex ib" FISCO
5. 1/2" NPT threads need to be sealed for dust and water ingress protection, IP 66, IP
FISCO parameters
B.7.4
N1 ATEX Type N
Certificate Nemko 10ATEX1072X
Standards EN 60079-0:2012, EN 60079-11:2012, EN 60079-15:2010, EN
Markings
Special Conditions for Safe Use (X):
1. The transmitter circuits does not withstand 500V AC dielectric strength test
Safety parameters HART 42.4 V
Safety parameters
Fieldbus
B.8

International

B.8.1
E7 IECEx Flameproof
Certificate IECEx NEM 06.0001X
Standards IEC 60079-0:2011, IEC 60079-1:2014-06, IEC 60079-11:2011; IEC
Markings
364
meatal surfaces. The area of the non-conducting part exceeds the maximum
permissible areas for Group III according to EN 60079-0.2012 clause 7.4:3.
Therefore, when the probe is used in a potentially explosive atmosphere group III,
EPL Da, appropriate measures must be taken to prevent electrostatic discharge.
power supply, when the power supply is certified with three separate safety current
limiting devices and voltage limitation which meets the requirements for type Ex ia.
67 or "Ex t", EPL Da or Db is required.
Ui
17.5 V
60079-31:2014
II 3G Ex nA ic IIC T4 Gc (-50 °C ≤ Ta ≤ +60 °C /+70 °C)
II 3G Ex ic IIC T4 Gc (-50 °C ≤ Ta ≤ +60 °C /+70 °C)
II 3D Ex tc IIIC T69 °C/T79 °C Dc (-50 °C ≤ Ta ≤ +60 °C /+70 °C)
according to EN 60079-11 clause 6.3.13 due to earth connected transient
suppressing devices. Appropriate measures have to be considered by installation.
Ui
32 V
60079-26:2014, IEC 60079-31:2013
Ex db ia IIC T4 Ga/Gb (-40 °C ≤ Ta ≤ +60 °C /+70 °C)
Ex ta IIIC T69 °C/T79 °C Da (-40 °C ≤ Ta ≤ +60 °C /+70 °C)
Um=250 VAC, IP66/IP67
Ii
Pi
380 mA
5.32 W
Ii
Pi
23 mA
1 W
21 mA
0.7 W
Rosemount 5300 Level Transmitter
Reference Manual
00809-0100-4530
Ci
Li
4.95 nF
<1 µH
Ci
Li
7.25 nF
Negligible
4.95 nF
Negligible

Hide quick links:

Advertisement

Table of Contents
loading