Agilent Technologies 1100 Series Reference Manual page 116

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector
The cell body can be fitted in two positions to allow the capillaries routed upwards or
NOTE
downwards (depending on where the column is located). Route the capillaries directly
column (inlet) and waste assembly (outlet).
NOTE
With the instrument accessory kit comes a 4-mm wrench and with the Sealing Kit a
special adapter. Both together work as a torque wrench with pre-defined torque (maximum allowed torque
116
1100 Series DAD and MWD Reference Manual