Setting The 8-Bit Configuration Switch - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

6
Introduction to the Detector

Setting the 8-bit Configuration Switch

The 8-bit configuration switch is located next to the GPIB connector. Switch settings
provide configuration parameters for GPIB address, serial communication protocol and
instrument specific initialization procedures.
Figure 66
Table 40
Switches 1 and 2 define which set of parameters (for example, for GPIB, RS-232C, and
so on) will be changed. Once the change has been completed, the instrument must be
powered up again in order to store the values in the non-volatile memory.
In the non-volatile memory, the parameters are kept, regardless of whether you turn the
instrument off and on again. They will be kept until the same set of parameters is
252
1100 Series DAD and MWD Reference Manual
Factory setting is
shown for the diode
array detector or
multiple wavelength
detector
8- bit Configuration Switch
8- bit Configuration Switch
Mode Select 1
GPIB
0
RS-232C
0
Reserved
1
TEST/BOOT
1
2
3
4
0
GPIB Address
1
Baud rate
0
Reserved
1
RSVD
SYS
5
6
7
Data
Parity
Bits
RSVD
RSVD
8
FC