Replacing The Holmium Oxide Filter Motor - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector

Replacing the Holmium Oxide Filter Motor

When required
Tools required
Parts required
NOTE
If the filter motor was removed, the filter lever should not be reused. Always use a new
filter lever to assure correct fit on the filter motor shaft.
Preparations for
this procedure:
Turn off the lamp.
Switch off the
detector, and
disconnect the
power cable.
Remove detector
from stack and
place it on working
bench.
Remove front
cover, top cover
and top foam
158
1100 Series DAD and MWD Reference Manual
If defective
Screwdriver POZI 1 PT3
hexagonal key 3 mm
Filter motor assembly G1315-68700 (includes filter lever G1315-45001
and
spring 1460-1510)
Hol iu O ide Filte
For others see
1
Reinstall the front cover, top cover and top foam
E ha gi g the Dete to Mai Boa d
section, see
on page 143.
Replace detector into the stack.
Reconnect the cables and turn on the detector.
on page 197
Place the optical unit on your bench.