If a set limit has been exceeded, a message box will pop up. This will not stop a sequence
or run (information only to plan maintenance activities). If you press Reset, the limits
will be removed. Ignore will continue to keep the EMF flag set.
Firmware Update
Use the Esc key to receive Views on the F5 key. Choose System from the pull-down
menu. Use the F3 key (Records) to select the DAD or MWD. Use the F5 key (FW
Update) to enter the Update section. If you want to update the resident firmware
(together with specific main firmware revisions), select the a file from the PCMCIA
card (RES_n_nn.DLB) and press execute. If you want to update the main firmware,
press F7 key (Transfer) to turn the module into the resident mode (LED on module
should blink yellow).
Use the Esc key to receive Views on the F5 key. Choose System from the pull-down
menu. Use the F3 key (Records) to select the Generic module. In this screen the
resident firmware revision is shown.