Agilent Technologies 1100 Series Reference Manual

Micro vacuum degasser
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Agilent 1100 Series
Micro Vacuum Degasser
Reference Manual
s1

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 1100 Series

  • Page 1 Agilent 1100 Series Micro Vacuum Degasser Reference Manual...
  • Page 2 85. Reproduction, adaption, change without notice. or translation without Warning Symbols Used prior written permission Agilent Technologies In This Book is prohibited, except as makes no warranty of allowed under the any kind with regard to copyright laws.
  • Page 3 Agilent 1100 Series Micro Vacuum Degasser Reference Manual...
  • Page 4 In This Book This manual contains technical reference information about the Agilent 1100 Series micro vacuum degasser. The manual describes the following: • installation, • optimizing performance, • troubleshooting, • repairing, • parts and materials, • theory of operation, and •...
  • Page 5: Table Of Contents

    Contents 1 Installing the Micro Vacuum Degasser How to install the micro vacuum degasser Site Requirements 10 Unpacking the Micro Vacuum Degasser 12 Optimizing the Stack Configuration 14 Installing the Micro Vacuum Degasser 16 Flow Connections to the Micro Vacuum Degasser 18 Transporting the Micro Vacuum Degasser 21 2 Optimizing Performance How to optimize the micro vacuum degasser to achieve best...
  • Page 6 Contents Hardware Symptoms All Lamps are Off 37 If the Status Indicator is Red 39 If the Status Indicator is Yellow and the Vacuum Pump is not Running 40 If the Status Lamp is Yellow and the Vacuum Pump Runs Continuously 41 4 Repairing the Micro Vacuum Degasser Instructions on how to repair the micro vacuum degasser...
  • Page 7 The vacuum degasser’s operation and electronics How the Electronics Work 74 7 Specifications Performance specifications of the micro vacuum degasser Performance Specifications 80 Legal, Safety and Warranty Information 81 Warranty Statement Safety Information Radio Interference Sound Emission Agilent Technologies on Internet...
  • Page 8 Contents...
  • Page 9: Installing The Micro Vacuum Degasser

    Installing the Micro Vacuum Degasser How to install the micro vacuum degasser...
  • Page 10: Site Requirements

    Shock hazard or damage of your instrumentation can result, if the devices are connected to a line voltage higher than specified. WAR NI N G Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
  • Page 11 Installing the Micro Vacuum Degasser Site Requirements If the bench should carry a complete Agilent 1100 Series system, make sure that the bench is designed to carry the weight of all the modules. Environment Your micro vacuum degasser will work within specifications at ambient temperatures and relative humidity as described in Table 1.
  • Page 12: Unpacking The Micro Vacuum Degasser

    Ensure all parts and materials have been delivered with the micro vacuum degasser. The delivery checklist is shown below. To aid in parts identification, please see Chapter 5 “Parts and Materials”. Please report missing or damaged parts to your local Agilent Technologies Sales and Service Office. Table 2...
  • Page 13 Installing the Micro Vacuum Degasser Unpacking the Micro Vacuum Degasser Accessory Kit Contents Table 3 Accessory Kit Contents G1322-68705 Description Part Number Quantity 5062-8534 Syringe Syringe Adapter 9301-1337 Mounting Tool 0100-1710 5062-2463 Waste Tube Connecting Tubes labeled A to D G1322-67300 Reorder Number (pack of 10) Reorder Number (5m)
  • Page 14: Optimizing The Stack Configuration

    Installing the Micro Vacuum Degasser Optimizing the Stack Configuration Optimizing the Stack Configuration If your micro vacuum degasser is part of a system, you can ensure optimum performance by limiting the configuration of the system stack to the following configuration. This configuration optimizes the system flow path, ensuring minimum delay volume (from point of solvent mixing to head of column) and dead volume (from point of injection to detector outlet).
  • Page 15 Installing the Micro Vacuum Degasser Optimizing the Stack Configuration Figure 2 Recommended Stack Configuration (Rear View) Remote cable Pressure output to recorder CAN Bus cable to handheld controller CAN Bus cable for inter AC power module communication Analog signal to recorder GPIB to LC ChemStation...
  • Page 16: Installing The Micro Vacuum Degasser

    N O T E In an Agilent 1100 Series system, the individual modules are connected through a CAN cable. The Agilent 1100 Series micro vacuum degasser is an exception. The micro vacuum degasser can be connected via the APG remote connector to the other modules of the stack.
  • Page 17 Installing the Micro Vacuum Degasser Installing the Micro Vacuum Degasser Figure 3 Rear of the Micro Vacuum Degasser Power socket Fuse holder Figure 4 Front of the Micro Vacuum Degasser Status indicator Power switch Waste outlet Serial number 5 Press in the power switch to turn on the micro vacuum degasser. N O T E The power switch stays pressed in and a green indicator lamp in the power switch is on when the micro vacuum degasser is turned on.
  • Page 18: Flow Connections To The Micro Vacuum Degasser

    2 Remove the front cover by pressing the snap fasteners on both sides. Figure 5 Removing the Front Cover 3 If the micro vacuum degasser is not used with an Agilent 1100 Series pump, connect the waste tube from the accessory kit to the waste outlet and place into your waste system.
  • Page 19 Installing the Micro Vacuum Degasser Flow Connections to the Micro Vacuum Degasser 5 Connect the solvent tubes from the bottle head assemblies to the inlet connectors A through D (typically the upper connection of the channel) of the micro vacuum degasser. Use the mounting tool shown in Figure 6 to fix the tube screw.
  • Page 20 Installing the Micro Vacuum Degasser Flow Connections to the Micro Vacuum Degasser Figure 7 Flow Connections to the Micro Vacuum Degasser Bottle Head Assembly Solvent cabinet Vacuum Degasser Inlet tube Outlet tube Tube clip Purge valve Pump head Inlet adapter Waste tubing Outlet capillary to sampler MCGV...
  • Page 21: Transporting The Micro Vacuum Degasser

    Installing the Micro Vacuum Degasser Transporting the Micro Vacuum Degasser Transporting the Micro Vacuum Degasser WAR NI N G Drain any remaining solvents from the degassing channels before transporting the micro vacuum degasser. 1 Disconnect the solvent tubes from solvent inlets from front panel. 2 Disconnect one of the solvent tubes from your pump.
  • Page 22 Installing the Micro Vacuum Degasser Transporting the Micro Vacuum Degasser...
  • Page 23: Optimizing Performance

    Optimizing Performance How to optimize the micro vacuum degasser to achieve best chromatographic results...
  • Page 24: Introduction To The Micro Vacuum Degasser

    Optimizing Performance Introduction to the Micro Vacuum Degasser Introduction to the Micro Vacuum Degasser The micro vacuum degasser comprises a 4-channel vacuum container, including 4 micro structured membranes, and a vacuum pump. When the micro vacuum degasser is switched on, the control circuit turns on the vacuum pump which generates a partial vacuum in the vacuum container.
  • Page 25: When To Use A Micro Vacuum Degasser

    Optimizing Performance When to Use a Micro Vacuum Degasser? When to Use a Micro Vacuum Degasser? For capillary LC application with our 1100 Series Capillary LC System For high-pressure mixing pumps with low flow rates or when you have following conditions: •...
  • Page 26: Operation Modes Of The Micro Vacuum Degasser

    Optimizing Performance Operation Modes of the Micro Vacuum Degasser Operation Modes of the Micro Vacuum Degasser The micro vacuum degasser allows various operation modes. Operation mode 1 and 2 are the normal operation modes of the degasser: • In operation mode 1 the vacuum degasser works at 115 Torr. •...
  • Page 27 Optimizing Performance Operation Modes of the Micro Vacuum Degasser N O T E The voltage readings given for the various modes are approximate values. All values are set in the factory according to the vacuum level. Depending on the vacuum sensor batch variations this will result in different voltage readings. Regardless of the readings of the various stages no adjustment should be performed on the degasser electronics.
  • Page 28 Optimizing Performance Operation Modes of the Micro Vacuum Degasser If the vacuum level of normal operation mode two cannot be achieved within 8 minutes the instrument turns into error mode. Continuous Mode The continuous mode is activated either by switching SW1 on the main board to 1 (removal of cover is required, see “Removing and Refitting the Top Cover”...
  • Page 29: Operational Hints For The Micro Vacuum Degasser

    Optimizing Performance Operational Hints for the Micro Vacuum Degasser Operational Hints for the Micro Vacuum Degasser Priming the Degasser The micro micro vacuum degasser can be primed either by drawing solvent through the degasser with a syringe or by pumping with the connected pump. Priming the micro micro vacuum degasser with a syringe is recommended, when: •...
  • Page 30 Optimizing Performance Operational Hints for the Micro Vacuum Degasser 3 Disconnect solvent outlet tube of the channel that is supposed to be primed from your pump. 4 Connect syringe adapter to solvent outlet tube. 5 Push syringe adapter onto syringe. 6 Pull syringe plunger to draw at least 5 ml of solvent through degasser and tubing.
  • Page 31: Solvent Information

    Optimizing Performance Solvent Information Solvent Information Always filter solvents through 0.4 µm filters, small particles can permanently block the capillaries. Avoid the use of the following steel-corrosive solvents: • Solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on).
  • Page 32: Prevent Blocking Of Solvent Filters

    Optimizing Performance Prevent Blocking of Solvent Filters Prevent Blocking of Solvent Filters Contaminated solvents or algae growth in the solvent bottle will reduce the lifetime of the solvent filter and will influence the performance of the pump. This is especially true for aqueous solvents or phosphate buffers (pH 4 to 7). The following suggestions will prolong lifetime of the solvent filter and will maintain the performance of the pump.
  • Page 33: Troubleshooting

    Troubleshooting Errors, their probable cause, and the action required to correct them...
  • Page 34: Status Indicators

    Status Indicators Two status lamps are located on the front of the micro vacuum degasser. The left lamp indicates the power supply status, the right lamp indicates the micro vacuum degasser status. Figure 9 Location of Status Lamps Status indicator Power supply lamp...
  • Page 35: Power Supply Lamp

    Troubleshooting Power Supply Lamp Power Supply Lamp The power supply lamp is integrated into the main power switch. When the lamp is illuminated (green), the power is ON. Instrument Status Indicator The instrument status indicator indicates one of three possible instrument conditions.
  • Page 36: Hardware Symptoms

    Vacuum Degasser” on page 16. The micro vacuum degasser itself will not be able to generate any error messages in the system logbook of the Agilent 1100 Series. The following pages describe hardware symptoms which help you to isolate the cause of a hardware failure.
  • Page 37: All Lamps Are Off

    Troubleshooting All Lamps are Off All Lamps are Off If all other modules in the system are on (power switch lamp is green) and are recognized by the connected user interface (module parameters can be set, module-specific screens appear, and so on), then do the following to determine the problem with the micro vacuum degasser: Ensure the power cable is connected to the degasser, and the power cable is connected to line power.
  • Page 38 Troubleshooting All Lamps are Off 5 Ensure the fuse wires inside the fuses are not broken. If a test meter is available, check the resistance of each fuse. A good fuse shows a low resistance (approximately 0 Ohm). 6 If a fuse is defective (wire broken or high resistance), insert a new fuse. 7 Reinsert the fuse holders and the power cable.
  • Page 39: If The Status Indicator Is Red

    Troubleshooting If the Status Indicator is Red If the Status Indicator is Red Sufficient vacuum is normally built up during the initial start-up and is maintained by turning on the vacuum pump whenever triggered by the vacuum sensor. If the vacuum cannot be reached, or is lost faster than expected during operation, the micro vacuum degasser will be forced into an error state.
  • Page 40: If The Status Indicator Is Yellow And The Vacuum Pump Is Not Running

    Troubleshooting If the Status Indicator is Yellow and the Vacuum Pump is not Running If the Status Indicator is Yellow and the Vacuum Pump is not Running Remove the top cover (see “Removing and Refitting the Top Cover” on page 50). Check the operation of the vacuum pump.
  • Page 41: If The Status Lamp Is Yellow And The Vacuum Pump Runs Continuously

    Troubleshooting If the Status Lamp is Yellow and the Vacuum Pump Runs Continuously If the Status Lamp is Yellow and the Vacuum Pump Runs Continuously Before starting the following troubleshooting procedures check that all tubes on the vacuum container, the pressure sensor, the vacuum pump and the solenoid valve are tight and connected properly.
  • Page 42 Troubleshooting If the Status Lamp is Yellow and the Vacuum Pump Runs Continuously Results A pressure reading below 500 mV indicates a good functioning vacuum pump. A pressure reading above 600 mV indicates that one of the pump chambers is contaminated or defective.
  • Page 43 Troubleshooting If the Status Lamp is Yellow and the Vacuum Pump Runs Continuously 6 Turn ON the degasser, wait 15 seconds, then connect the solenoid valve tube at position 9. Results A pressure reading below 500 mV indicates a good working vacuum chamber 1.
  • Page 44 Troubleshooting If the Status Lamp is Yellow and the Vacuum Pump Runs Continuously Figure 11 Block Diagram Control Electronics Switch Leak tub Solenoid Valve Sensor Assembly Vacuum Vacuum chamber chamber Vacuum Pump Leak Leak...
  • Page 45: Repairing The Micro Vacuum Degasser

    Repairing the Micro Vacuum Degasser Instructions on how to repair the micro vacuum degasser...
  • Page 46 Repairing the Micro Vacuum Degasser Figure 12 Overview of Repair Procedures Control Assembly and Sensor Assembly, see page 60 Vacuum Chambers, see page 55 Solenoid Valve, see page 59 Vacuum Pump, see page 57...
  • Page 47 Repairing the Micro Vacuum Degasser WAR NI N G The following procedures require opening the main cover of the micro vacuum degasser. To prevent personal injury, the power cable must be removed from the micro vacuum degasser before opening the cover. Do not connect the power cable to the micro vacuum degasser while the covers are removed.
  • Page 48: Cleaning The Instrument

    Repairing the Micro Vacuum Degasser Cleaning the Instrument Cleaning the Instrument The micro vacuum degasser case should be kept clean. Cleaning should be done with a soft cloth slightly dampened with water or a solution of water and a mild detergent. Do not use an excessively damp cloth that liquid can drip into the micro vacuum degasser.
  • Page 49: Using The Esd Strap

    Repairing the Micro Vacuum Degasser Using the ESD Strap Using the ESD Strap Electronic boards are sensitive to electrostatic discharge (ESD). In order to prevent damage, always use an ESD strap supplied with the standard accessory kit when handling electronic boards and components. Using the ESD Strap 1 Unwrap the first two folds of the band and wrap the exposed adhesive side firmly around your wrist.
  • Page 50: Removing And Refitting The Top Cover

    Repairing the Micro Vacuum Degasser Removing and Refitting the Top Cover Removing and Refitting the Top Cover WAR NI N G When opening capillary or tube fittings solvents may leak out. Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet by the solvent vendor, especially when toxic or hazardous solvents are used.
  • Page 51 Repairing the Micro Vacuum Degasser Removing and Refitting the Top Cover 4 Unscrew the screws on the top plate and 5 Place the metal cover onto the housing. remove the plate by lifting its back and then Lower the front of the plate first, then slide sliding the plate to the front.
  • Page 52: Assembling The Main Cover

    Repairing the Micro Vacuum Degasser Assembling the Main Cover Assembling the Main Cover When required If cover is broken. Tools required None Parts required Cover kit 5062-8579 (includes base, top, left and right) N O T E The cover kit contains all parts, but it is not assembled. WAR NI N G In case you insert the left or right side in the opposite position, you may not be able to remove the side from the top part.
  • Page 53: Exchanging The Status Light Pipe

    Repairing the Micro Vacuum Degasser Exchanging the Status Light Pipe Exchanging the Status Light Pipe When required If part is broken Tools required Screwdriver Pozidriv#1 Part required Status light pipe 5041-8384 Preparation for this procedure: 1 The status light pipe is clipped into the top cover.
  • Page 54: Overview Of Internal Parts

    Overview of Internal Parts Figure 14 shows a top view of the micro vacuum degasser. It shows the connections of the vacuum tubes as well as the electrical connections at the control assembly. Figure 14 Top View of the Micro Vacuum Degasser Control assembly Board clip Board clip...
  • Page 55: Exchanging A Vacuum Chamber

    Repairing the Micro Vacuum Degasser Exchanging a Vacuum Chamber Exchanging a Vacuum Chamber When required If internal membrane defective or vacuum chamber damaged Tools required Screwdriver Pozidriv #1 Material Vacuum chamber, G1379-60001 Preparation Disconnect the solvent inlet tubes at the degasser inlet ports Remove the solvent cabinet from the micro vacuum degasser WAR NI N G Drain any remaining solvents from the degassing channels before...
  • Page 56 Repairing the Micro Vacuum Degasser Exchanging a Vacuum Chamber Figure 15 Exchanging a Vacuum Chamber Vacuum chambers...
  • Page 57: Exchanging The Vacuum Pump

    Repairing the Micro Vacuum Degasser Exchanging the Vacuum Pump Exchanging the Vacuum Pump When required If not generating sufficient vacuum or electrically defective Tools required Screwdriver Pozidriv #1 Part required Vacuum pump, G1322-60000 1 Remove Top Cover, see “Removing and Refitting the Top Cover” on page 50. 2 Disconnect vacuum pump cable from connector CN2 at the control assembly (for identifying internal parts see “Overview of Internal Parts”...
  • Page 58 Repairing the Micro Vacuum Degasser Exchanging the Vacuum Pump Figure 16 Exchanging the Vacuum Pump Screws Fixing plate Vacuum pump Rubber foot and holder...
  • Page 59: Exchanging The Solenoid Valve

    Repairing the Micro Vacuum Degasser Exchanging the Solenoid Valve Exchanging the Solenoid Valve When required Valve not switching correctly or leaking Tools required Screwdriver Pozidriv #1 Part required Solenoid valve G1322-60003 1 Remove Top Cover, see “Removing and Refitting the Top Cover” on page 50. 2 Disconnect the solenoid connector from the control board (CN3) (for identifying internal parts see “Overview of Internal Parts”...
  • Page 60: Exchanging The Degasser Control Assembly And The Sensor Assembly

    Repairing the Micro Vacuum Degasser Exchanging the Degasser Control Assembly and the Sensor Assembly Exchanging the Degasser Control Assembly and the Sensor Assembly When required No power available, vacuum pump or switching valve not activated Tools required Screwdriver Pozidriv #1 Part required Degasser control assembly (includes the sensor assembly) G1322-66500 WAR NI N G...
  • Page 61 Repairing the Micro Vacuum Degasser Exchanging the Degasser Control Assembly and the Sensor Assembly 11 Clip the power switch light pipe to the degasser control assembly power switch. 12 Replace the two holding clips and fix the holding screws. 13 Connect all the cables to the control board (sensor assembly cable to CN6, solenoid valve cable to CN3, pump cable to CN2).
  • Page 62 Repairing the Micro Vacuum Degasser Exchanging the Degasser Control Assembly and the Sensor Assembly...
  • Page 63: Parts And Materials

    Parts and Materials Detailed illustrations and listings for parts and materials identification...
  • Page 64: Micro Vacuum Degasser Parts

    Parts and Materials Micro Vacuum Degasser Parts Micro Vacuum Degasser Parts Figure 19 Micro Vacuum Degasser Parts...
  • Page 65 Parts and Materials Micro Vacuum Degasser Parts Table 5 Micro Vacuum Degasser Parts Item Description Part Number Vacuum chamber G1379-60001 Sensor assembly (included in the control assembly) no PN Vacuum degasser control assembly G1322-66500 Fuse 500 mA 2110-0458 Board clip G1322-43100 Solenoid valve G1322-60003...
  • Page 66: Cover Parts

    Parts and Materials Cover Parts Cover Parts Table 6 Cover Parts Item Description Part Number Set of all plastic covers, top-side-base 5062-8579 Front cover 5062-8580 Logo plate, Agilent 1100 5042-1384 Tube clip 5041-8387 Figure 20 Cover Parts...
  • Page 67: Sheet Metal Kit

    Parts and Materials Sheet Metal Kit Sheet Metal Kit Table 7 Sheet Metal Kit Item Description Part Number Sheet metal kit, includes base and top plate G1379-68701 Figure 21 Sheet Metal Kit...
  • Page 68: Power And Status Light Pipes

    Parts and Materials Power and Status Light Pipes Power and Status Light Pipes Table 8 Power and Status Light Pipes Item Description Part Number Power switch coupler 5041-8383 Light pipe – power switch 5041-8382 Power switch button 5041-8381 Light pipe – status lamp 5041-8384 Figure 22 Power and Status Light Pipes...
  • Page 69: Accessory Kit

    Parts and Materials Accessory Kit Accessory Kit Table 9 Accessory Kit G1322-68705 Item Description Part Number 5062-8534 Syringe Syringe adapter 9301-1337 Solvent tubing kit (4 tubes degasser to pump) G1322-67300 Waste tube** 5062-2463 Mounting Tool 0100-1710 Reorder number (pack of 10) **Reorder number (5m)
  • Page 70: Cable Overview

    Cable Overview WAR NI N G Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations. Table 10 Cable Overview Type Description Part Number Remote Agilent 1100 modules...
  • Page 71: Remote Cables

    Remote Cables One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent 1100 Series modules. The other end depends on the instrument to be connected to. Agilent 1100 to HP 1050, HP 1046A or 35900 A/D Converters...
  • Page 72: Auxiliary Cable

    PIN 6 One end of this cable provides a modular plug to be connected to the vacuum degasser. The other end is for general purpose. Agilent 1100 Series Vacuum Degasser to General Purpose Connector G1322-61600 Color Pin Agilent 1100 Signal Name...
  • Page 73: Theory Of Operation

    Theory of Operation The vacuum degasser’s operation and electronics...
  • Page 74: How The Electronics Work

    Theory of Operation How the Electronics Work How the Electronics Work The vacuum degasser has two different normal operation modes and a continuous mode. In operation mode 1 the vacuum degasser works around a defined set point (115 Torr). Due to environmental conditions it is possible that the vacuum degasser cannot reach the pre-defined set point.
  • Page 75 Theory of Operation How the Electronics Work The pressure sensor is connected to the vacuum chamber and checks for the correct vacuum in the system. The amplifier and comparator circuit determines the working range of the vacuum that has to be built up. When the vacuum degasser is turned on and the vacuum in the chamber is not within working range (above error limit of 190 Torr), the amplifier and comparator circuit sends a signal to the vacuum pump driver and the timers of the vacuum pump (timer 1) and the solenoid...
  • Page 76 Theory of Operation How the Electronics Work and solenoid valve. The status lamp turns red and the error output on the remote connector will be activated. The error output at the remote connector provides a contact closure (potential free open collector maximum load 35 V DC/50 mA) as long as the error condition is active.
  • Page 77 Theory of Operation How the Electronics Work Figure 24 Block Diagram — Vacuum Degasser Control Assembly Vacuum Inverter Timer 1 Driver pump Continuous Mode Solenoid Timer 2 Driver valve Timer Mode Timer 3 Amplifier Pressure Red LED Error Monitor Error timer Error driver &...
  • Page 78 Theory of Operation How the Electronics Work...
  • Page 79: Specifications

    Specifications Performance specifications of the micro vacuum degasser...
  • Page 80: Performance Specifications

    For pressure monitoring, range 0 – 3 V N O T E The Agilent 1100 Series micro vacuum degasser has been tested for evaporation of solvents into the atmosphere by an independent institute with approved methods. The tests were performed with Methanol (BIA Nr. 7810) and Acetonitrile (NIOSH, Nr.
  • Page 81: Legal, Safety And Warranty Information

    Legal, Safety and Warranty Information...
  • Page 82: Warranty Statement

    Warranty Statement Warranty Statement All Chemical Analysis Products Agilent Technologies warrants its chemical analysis products against defects in materials and workmanship. For details of the warranty period in your country, call Agilent. During the warranty period, Agilent will, at its option, repair or replace products which prove to be defective.
  • Page 83 Warranty Statement 4 operation outside of the environmental and electrical specifications for the product, 5 improper site preparation and maintenance, or 6 customer induced contamination or leaks. THE WARRANTY SET FORTH IS EXCLUSIVE AND NO OTHER WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED. AGILENT SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
  • Page 84 Warranty Statement Responsibilities of Agilent Technologies Agilent Technologies will provide warranty services as described in Table 12. Table 12 Warranty Services Services During Warranty Warranty Period Type Agilent 1100 Series of Modules 1 Year Onsite GC, LC, UV-Visible, and LAS supplies and...
  • Page 85: Safety Information

    Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. General...
  • Page 86 Safety Information Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. Do not install substitute parts or make any unauthorized modification to the instrument.
  • Page 87 Safety Information WAR NI N G A warning alerts you to situations that could cause physical injury or damage to the equipment. Do not proceed beyond a warning until you have fully understood and met the indicated conditions. CA UT IO N A caution alerts you to situations that could cause a possible loss of data.
  • Page 88: Radio Interference

    Radio Interference Radio Interference Manufacturer’s Declaration This is to certify that this equipment is in accordance with the Radio Interference Requirements of Directive FTZ 1046/1984. The German Bundespost was notified that this equipment was put into circulation, the right to check the series for compliance with the requirements was granted. Test and Measurement If test and measurement equipment is operated with equipment unscreened cables and/or used for measurements on open set-ups, the user has to assure...
  • Page 89: Agilent Technologies On Internet

    Agilent Technologies on Internet Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com/chem...
  • Page 90 Agilent Technologies on Internet...
  • Page 91 Index exchanging maximum sensitivity, 25 accessory kit, 13, 49, 69 control asssembly, 60 mounting tool, 13, 19 Agilent on internet, 89 sensor assembly, 60 algae growth, 32 solenoid valve, 59 ambient non-operating temperature, 11 status light pipe, 53 non-operating altitude, 11 ambient operating temperature, 11 vacuum chamber, 55 number of channels, 80...
  • Page 92 Index shipping container, 12 site requirements, 10 snap fastener, 18 solenoid valve, 59 solvent bottle, 18 solvent cabinet, 18, 50 solvent filters checking, 32 cleaning, 32 prevent blocking, 32 solvent information, 26 solvent outlet tubes, 18 specification performance specifications, 80 physical, 11 stack configuration, 14 status indicator, 34, 39, 40, 41...
  • Page 94 In This Book This manual contains technical reference information about the Agilent 1100 Series micro vacuum degasser. The manual describes the following: • installation, • optimizing performance, • troubleshooting, • repairing, • parts and materials, • theory of operation, and...

Table of Contents