Agilent Technologies 1100 Series Reference Manual page 152

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector
If the filter motor was removed, the filter lever should not be reused. Always use a new
NOTE
filter lever to assure correct fit on the filter motor shaft.
Preparations for
this procedure:
Turn off the lamp.
Switch off detector
and disconnect the
power cable.
Remove detector
from stack and
place it on the
working bench.
Remove the front
cover, top cover
and top foam
section, see
Re o i g the Top
Co e a d Foa
on page 140.
Remove the optical
unit, see
Re o i g the
Optical Unit
G
A/G
A
on page 154 or
Re o i g the
Optical Unit
G
B/G
B
on page 156.
152
1100 Series DAD and MWD Reference Manual
Spectro window 79880-28111
Filter lever G1315-45001
1
Place the optical unit on your bench.