Electronic boards and components are sensitive to electronic discharge (ESD). In order
CAUTION
to prevent damage always use an ESD protection (for example, the ESD wrist strap from
the accessory kit) when handling electronic boards and components, see
98.
1 Turn off the lamp.
2 Switch off the detector and disconnect the cables.
3 Remove detector from stack and place it on the working bench.
4 Remove the front cover, top cover and top foam section, see
Cover and Foam"
5 Remove the optical unit, see
page 154 or
6 Disconnect the leak sensor cable from the detector main board (J21).
On the G1315B DAD and G1365B MWD the leak sensor can be pulled out without the
NOTE
removing the leak pan (the next steps describe the procedure for the G1315A/G1365A).
12 Remove the leak pan from the cabinet bottom.
1100 Series DAD and MWD Reference Manual
on page 140.
"Removing the Optical Unit (G1315A/G1365A)"
"Removing the Optical Unit (G1315B/G1365B)"
165
Repairing the Detector
Usi g the ESD St ap
on page
"Removing the Top
on
on page 156.