Agilent Technologies 1100 Series Reference Manual page 32

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
How to optmize the Detector
2
Connection of flow cell
For flow rates from 0.5 ml/min connect column using the zero-
dead-volume fittings of the detector.
For small column i.d. (e.g 1 mm) the inlet capillary of the micro
flow cell can be connected directly to the column.
3
Setting the peak width (response time)
Use peak width according
Set the peak-width close to the width of a narrow peak of interest
in your chromatogram.
4 Setting wavelength and bandwidth
• Sample wavelength:
Never miss a peak by the use of a browser wavelength like 250 nm
with 100 nm bandwidth.
Select specific wavelength with reduced bandwidth if you need
selectivity, e.g. 250,10 nm and 360,100 nm as reference
wavelength.
Set the sample wavelength to a peak or valley in the spectrum to
get best linearity for high concentrations.
Reference wavelength:
Select the reference wavelength with broad bandwidth (30...100
nm) wavelength range where your analytes have little or no
absorbance (e.g. sample at 254 nm, reference at 320 nm).
Table 4 Optimization Overview, continued
Parameter
5
Setting the slit width
Use 4 nm slit for normal applications.
Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance
bands and for high concentrations.
Use a wide slit (e.g. 16 nm) to detect very low concentrations.
Optimizing spectral acquisition (DAD only)
Select spectra acquisition mode according to your needs (see
Optio s fo sto i g spe t a
Set the spectral wavelength range (for colorless samples 190...400
nm is sufficient).
Set step to 4 nm for normal use; set small step (and slit width) if
high resolution of spectra with fine structure is wanted.
Typical column
Typical peak
length
width
T<= 5 cm
0.025 min
32
1100 Series DAD and MWD Reference Manual
Figure 7
as starting point.
on page 42).
Recommended flow cell
chromatographic resolution
• peak resolution versus sensitivity versus
disk space
• sensitivity versus selectivity
• sensitivity versus linearity
• baseline drift due to RI effects.
Impact
spectral resolution, sensitivity and
linearity.
High pressure
flow cell for

Hide quick links:

Advertisement

Table of Contents