Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Figure 23
Intensity Test Evaluation
The Aligent ChemStation evaluates four spectral ranges automatically, and displays the
limits for each range, the measured intensity counts, and passed or failed for each
spectral range (see Figure 24).
Figure 24
Test Failed
• Absorbing solvent or air bubble in flow cell.
• Dirty or contaminated flow cell.
• Dirty or contaminated optical components (achromat, windows).
• Old or non-Agilent lamp.
1100 Series DAD and MWD Reference Manual
Intensity Test
Intensity Test Results
Probable Causes
Troubleshooting and Test Functions
79