Agilent Technologies 1100 Series Reference Manual page 148

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector
Use this procedure for the G1315A/G1365A only.
NOTE
1 Turn off the lamp.
2 Switch off detector, and disconnect power cable.
3 Remove detector from stack and place it on working bench.
4 Remove front cover, top cover and top foam section, see
5 Remove both lamps from the optical unit, see
6 Remove flow cell, see
7 Remove flow cell cover, see
148
1100 Series DAD and MWD Reference Manual
and Foam"
on page 140.
(G1315A/G1365A only!).
page 127 (G1315A/G1365A only!).
"Exchanging a Flow Cell"
"Cleaning or Exchanging the Holmium Oxide Filter"
"Removing the Top Cover
"Exchanging a Lamp"
on page 105.
on page 102
on