Agilent Technologies 1100 Series Reference Manual page 257

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

again resets the power supply to normal operation if the cause of the overload condition
has been removed.
An over-temperature sensor in the main power supply is used to turn off output voltages
if the temperature exceeds the acceptable limit (for example, if the cooling fan of the
instrument fails). To reset the main power supply to normal operating conditions, turn
the instrument off, wait until it is approximately at ambient temperature and turn the
instrument on again.
The following table gives the specifications of the main power supply.
Table 48
Maximum power
Line Input
Output 1
Output 2
1100 Series DAD and MWD Reference Manual
Main Power Supply Specifications
300 VA / 130 W
100 – 120 or 220 – 240 volts AC
± 10 %, line frequency of 50/60 Hz
+ 24 V / 4.5 A (maximum)
+ 36 V / 2.5 A (maximum)
257
Introduction to the Detector
Continuous output
Wide ranging
total power consumption of + 24 V
and + 36 V must not exceed 107 W.
6

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents