Declaration of Conformity for HOX2 Filter

We herewith inform you that the holmium oxide glass filter (type Hoya HY-1)
absorbance maxima positions:
Agilent Technologies guarantees the traceability of the specified absorbance maxima to
a National Institute of Standards & Technology (NIST) holmium oxide solution
standard with a lot-to-lot tolerance of ±0.3nm.
The wavelength calibration filter built into the Agilent Technologies UV-visible
detectors Table 54
Product No.
is made of this material and meets these specifications. It is, therefore, suitable for
wavelength calibration of these detectors within the specified wavelength accuracy of
the respective detector over its wavelength range.
1100 Series DAD and MWD Reference Manual
( Part No. 79880-22711) meets the following specification of
361.0 nm - 418.9 nm - 453.7 nm - 536.7 nm (spectral bandwidth: 2 nm ).
79880A
79883A
79854A
1306A
G1315A/B
G1365A/B
G1600A
HP 1090 system
HP 1090 system
HP 1050 series
HP 1050 series
Agilent 1100 series
Agilent 1100 series
295
Safety Information
Wavelength Accuracy
± 1 nm
± 1 nm
± 1 nm
± 1 nm
± 1 nm
± 1 nm
± 1 nm