Replacing Leak Handling System Parts - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

1 Remove the front cover.
2 Use tissue to dry the leak sensor area and the leak pan.
3 Observe the capillary connections and the flow cell area for leaks and correct, if
4 Replace the front cover.
Figure 53

Replacing Leak Handling System Parts

When required
Tools required
Parts required
1 Remove the front cover.
2 Pull the leak funnel out of the leak funnel holder.
1100 Series FD Reference Manual 129
required.
Observing for Leaks
If the parts are corroded or
broken
None
Leak funnel 5061-3356
Leak funnel holder 5041-8389
Leak tubing (120 mm) 0890-
1711
Repairing the Fluorescence Detector