Exchanging The Lamp - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
Carefully remove the rear foam.

Exchanging the Lamp

When required
Tools required
Parts required
Wait until the lamp has cooled down before touching the lamp and/or triggerpack assembly.
WARNING
The instrument should not be operated without Xenon flash lamp. Otherwise
the Triggerpack electronics will be damaged.
NOTE
When loosening or removing screws take care that they do not fall into the instrument.
This may result in a complete removal of other assemblies.
1100 Series FD Reference Manual 135
A
Lamp: if noise or intensity exceeds limits or lamp does not
ignite
Screwdriver POZI 1 PT3
Hexagonal screwdriver 2.5 mm, 100 mm long
Hexagonal screwdriver 4mm, 100 mm long
Flash lamp 2140-0600
Repairing the Fluorescence Detector