Agilent Technologies 1100 Series Reference Manual page 52

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
First Steps with the Fluorescence Detector
limits, the purity factor is the mean purity value of all spectra that are within the purity
limits.
The reliability of the purity and the match factor depends on the quality of spectra
recorded. Because of the lower number of data points available with the fluorescence
detector in general, the match factors and purity data obtained show stronger deviations
compared to data from the diode array detector, even if the compounds are identical.
Table 11
reference sample.
Table 11 Peak confirmation using a fluorescence spectral library
Meas. Library CalTbl
RetTime
[min]
[min]
[min]
4.859
4.800
5.178
6.764
7.000
7.162
7.137
7.100
7.544
8.005
8.000
8.453
8.841
8.800
9.328
9.838
10.000
10.353
10.439
10.400
10.988
12.826
12.800
13.469
13.340
13.300
14.022
15.274
15.200
16.052
16.187
16.200
17.052
16.865
16.900
17.804
52
1100 Series FD Reference Manual
shows an automated library search based on the emission spectra from a PNA
Signal
Amount
[ng]
1
1.47986e-1
1
2.16156e-1
1
1.14864e-1
1
2.56635e-1
1
1.76064e-1
1
2.15360e-1
1
8.00754e-2
1
1.40764e-1
1
1.14082e-1
1
6.90434e-1
1
5.61791e-1
1
5.58070e-1
Purity
#
Match
Factor
-
1
993
-
1
998
-
1
995
-
1
969
-
1
993
-
1
997
-
1
1000
-
1
998
-
1
999
-
1
999
-
1
998
-
1
999
Libary Name
Naphthalene@em
Acenaphthene@em
Fluorene@em
Phenanthrene@em
Anthracene@em
Fluoranthene@em
Pyrene@em
Benz(a)anthracene@em
Chrysene@em
Benzo(b)fluoranthene@em
Benzo(k)fluoranthene@em
Benz(a)pyrene@em