Flash Trigger Lost - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Flash Trigger Lost

This message is displayed when the flash trigger is no longer generated.
Probable Causes
• firmware problem
• defective encoder
Suggested Actions
1100 Series FD Reference Manual 93
Reboot the detector (power cycle).
Multi Mode Off: exchange the firmware or replace the main board.
Multi Mode Ex: replace EX monochromator assembly.
Multi Mode Em: replace EM monochromator assembly.
Troubleshooting and Test Functions
4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1321a fld

Table of Contents