Agilent Technologies 1100 Series Reference Manual page 315

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Reference System:
Timetable
programing:
Table 59
Performance Specifications Agilent 1100 Series Fluorescence Detector,
Type
Spectrum acquisition:
Wavelength
characteristic
Flow cells
Control and data
evaluation
Analog outputs
Communications
in-line excitation measurement
up to 4 signal wavelengths,
response time, PMT Gain,
baseline behavior (append, free,
zero), spectral parameters
Specification
Comments
Excitation or Emission spectra
Scan speed: 28 ms per datapoint
(e.g. 0.6 s/spectrum 200-400 nm, 10 nm step)
Step size: 1-20 nm
Spectra storage: All
Repeatability+/- 0.2 nm
Accuracy+/- 3 nm setting
Standard: 8 µl volume and 20 bar
(2 MPa) pressure maximum,
quartz
Optional:
Fluorescence cuvette for offline
spectroscopic measurements
with 1 ml syringe, 8 µl volume,
quartz
Agilent ChemStation for LC,
Agilent 1100 Control Module
with limited spectral data analysis
and printing of spectra
Recorder/integrator: 100 mV or
1 V, output range >10
2
luminescence units, two outputs
Safety
and
Specifications
10
315