Wait until the lamp has cooled down before touching the lamp and/or triggerpack assembly.
The instrument should not be operated without Xenon flash lamp. Otherwise
the Triggerpack electronics will be damaged.
NOTE
When loosening or removing screws take care that they do not fall into the instrument.
This may result in a complete removal of other assemblies.
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell or disconnect the
tubings from the flow cell.
Remove detector from stack and place
it on the workbench.
Remove the covers as described in
Re o i g the Co ers
1100 Series FD Reference Manual 143
Tools required
Parts required
1
on page 146.
Repairing the Fluorescence Detector
Screwdriver POZI 1 PT3
Hexagonal screwdriver 2.5 mm, 100 mm
long
Hexagonal screwdriver 4 mm, 100 mm
long
EX condenser Assembly 1000-1136
Disconnect the emission and the
excitation cables and from the
detector main board.
5
B