Agilent Technologies 1100 Series Reference Manual page 121

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

strap from the accessory kit) when handling electronic boards and components (see
Strap Electronic boards are sensitive to electronic discharge (ESD). In order to prevent damage, always
use a ESD strap supplied i the sta dard a essor kit see A essor Kit o page
ha dli g ele tro i
There is a risk of damaging hardware due to overheating when operating the instrument without
covers.
CAUTION
Eye damage may result from directly viewing the light produced by the Xenon flash
WARNING
lamp used in this product. Always turn the xenon flash lamp off before removing it.
Cleaning the Detector
The detector case should be kept clean. Cleaning should be done with a soft cloth
slightly dampened with water or a solution of water and mild detergent. Do not use an
excessively damp cloth allowing liquid to drip into the detector.
WARNING
Do not let liquid drip into the detector. It could cause shock hazard and it could
damage the detector.
Using the ESD Strap
(ESD). In order to prevent damage, always use an ESD strap supplied in the standard
accessory kit (see
components.
1 Unwrap the first two folds of the band and wrap the exposed adhesive side firmly
2 Unroll the rest of the band and peel the liner from the copper foil at the opposite end.
3 Attach the copper foil to a convenient and exposed electrical ground.
1100 Series FD Reference Manual 121
oards a d o po e ts.
"Accessory Kit"
around your wrist.
Repairing the Fluorescence Detector
on page 133).
Electronic boards are sensitive to electronic discharge
on page 216) when handling electronic boards and
5
Usi g the ESD
6
he