Exchanging The Power Supply - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Reinstall the optical unit as described in
Enter the replacement in the maintenance logbook.
Reset the monochromator settings to default using the diagnose functions.
I this ase Wavelength Calibration Lost" is displayed and the instrument stays in Not Ready condition until you
perform a wavelength calibration.
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in
Wa ele gth Verifi atio a d Cali ratio
chapter

Exchanging the Power Supply

NOTE
The repair of the power supply assembly comprises exchanging the complete
assembly. No serviceable parts are inside.
1100 Series FD Reference Manual 191
I stalli g the Opti al U it
on page 105.
When required
If defective
Tools required
Screwdriver POZI 1 PT3
Wrench 1/4 inch
Wrench 5 mm
Wrench 7 mm
Parts required
Power supply 0950-
2528
Repairing the Fluorescence Detector
on page 194.
5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1321a fld

Table of Contents