Troubleshooting and Test Functions
7 Add the filter to the syringe and fit the needle to filter.
Figure 40
8 Lift the needle tip and carefully eject approximately 0.5 ml to remove air out of the
9 Add the PEEK fitting to the needle tip and fix both at the flow cell inlet.
NOTE
Do not inject the calibration sample without the sample filter.
10 Slowly inject about 0.2 ml and wait for about 10 seconds to inject another

Wavelength Calibration

1 From the user interface start the FLD Wavelength Calibration (see
If the wavelength calibration process fails, refer to
NOTE
page 101.
2 If a deviation is displayed, press Adjust and OK. The history table will be updated.
NOTE
106
1100 Series FD Reference Manual
Syringe with Sample Filter
syringe and to flush the needle.
0.1 ml. This will assure that the cell is filled properly.
Agilent ChemStation: Diagnosis - Maintenance - FLD Calibration Control
Module: System - Tests - FLD - Calibrate.
Figure
Wa ele gth Cali ratio Failed
41).
on