Replacing Pmt And/Or Flf Board - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Repairing the Fluorescence Detector
Next Steps:
Depending on the replacement you want to perform, proceed to the following procedures.

Replacing PMT and/or FLF board

When required
Tools required
Parts required
CAUTION
176
1100 Series FD Reference Manual
If defective or if application specific PMT is
required.
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm and 2.5 mm
Tweezers
Photo-multiplier Tube (PMT) 1970-0201
Photo-multiplier Tube (PMT) for other application ranges, please contact
Hamamatsu dealers. Alternative PMTs are either R928HA (185 to 900 nm)
or R3788HA (185 to 750 nm). No other PMTs are recommended.
FLF Board G1321-69531 (requires firmware revision A.04.06 or
above), FLR Reference Diode board G1321-66533