Replacing Leak Handling System Parts

When
If the parts are corroded or broken
Parts required
#
1
1
1
1 Remove the front cover.
2 Pull the leak funnel out of the leak funnel holder.
3 Pull out the leak funnel with the tubing.
4 Insert the leak funnel with the tubing in its position.
5 Insert the leak funnel into the leak funnel holder.
6 Replace the front cover.
Leak funnel
Leak funnel holder
Leak tubing
Agilent 1260 FLD User Manual
p/n
Description
5041-8389
Leak funnel
5061-3356
Leak funnel holder
5042-9974
Leak tubing (1.5 m, 120 mm required)
Figure 78

Replacing Leak Handling System Parts

Replacing Leak Handling System Parts
Leak sensor
9
Maintenance
193