UV-Radiation (UV-lamps only)

Emissions of ultraviolet radiation (200-315 nm) from this product is limited such that
radiant exposure incident upon the unprotected skin or eye of operator or service
personnel is limited to the following TLVs (Threshold Limit Values) according to the
American Conference of Governmental Industrial Hygienists:
Table 61
Exposure/day
8 hours
10 minutes
Typically the radiation values are much smaller than these limits:
Table 62
Position
Lamp installed, 50-cm distance
Lamp installed, 50-cm distance

Solvent Information

Observe the following recommendations on the use of solvents.
Flow Cell
1100 Series FD Reference Manual 325
UV-Radiation Limits
Effective Irradiance
2
0.1 µW/cm
2
5.0 µW/cm
UV-Radiation Typical Values
Effective Irradiance
2
average 0.016 µW/cm
2
maximum 0.14 µW/cm
Safety Information