Leak Sensor Short - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Troubleshooting and Test Functions

Leak Sensor Short

The leak sensor in the detector has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when
solvent cools the leak sensor, causing the leak sensor current to change within defined
limits. If the current increases above the upper limit, the error message is generated.
Probable Causes
• Defective leak sensor.
Suggested Actions
82
1100 Series FD Reference Manual
Exchange the leak sensor.