Agilent Technologies 1100 Series Reference Manual page 103

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Figure 39
12 You can now select a stored spectra and evaluate/print them. The maximum should
NOTE
The spectra on the control module is a real raw data spectrum. Therefore it looks not that
smooth as on the Agilent ChemStation (uses the spline algorithm).
NOTE
If the limits are not met, check for
The Wa ele gth Cali ratio Pro edure
Method Settings
Table 20 Method Settings
Setting
Peak Width
Fit Spectral Range
PMT Gain
Flash Lamp
Spectrum Range
Store Spectra
EX Wavelength
1100 Series FD Reference Manual 103
Evaluating Online Excitation Spectra
±
be found at 350 nm
3 nm.
on page 114.
Troubleshooting and Test Functions
I terpretatio of the Results
Value in WLEXTEST
>0.2 min (4 s, standard)
OFF
12
ON
EX 377 - 417 nm step 1 nm
All w/o signal
350 nm, ON
on page 125 or perform
Value in WLEMTEST
>0.2 min (4 s, standard )
OFF
12
ON
EM 330 - 370 nm step 1 nm
All w/o signal
350 nm, OFF
4