4
When the fan is inserted assure
that the air flow is from the rear
towards the front of the detector.
Insert the fan.
Lift the main board slightly.
Reconnect the fan cable and route it
underneath the board.

Exchanging the Leak Sensor or Leak Panel

When required
Tools required
Parts required
1100 Series FD Reference Manual 157
5
Carefully insert the rear foam. Reconnect the monochromator
Repairing the Fluorescence Detector
cables to the main board. Left is excitation, right is emission.
Rear
foam
A
EM
EX
B
If defective
Screwdriver POZI 1 PT3
Flat screwdriver
Leak sensor assembly 5061-
3356
Leak Pan G1321-45501
A
5
B