SATO CL4NX Service Manual page 45

Hide thumbs Also See for CL4NX:
Table of Contents

Advertisement

6
Press the
/
buttons to change the
Emit value until the Sensor value is below
0.5 (V). Set the Emit value as low as
possible.
7
If the Sensor value does not decrease
below 0.5 after you changed the Emit
value, press the
the Receive value.
8
Take a note of the Sensor value from the
above procedure. This is the "Low" level
value for the Gap sensor.
Next, check the "High" level (voltage) of the Gap sensor as follows:
9
Pass the media . (if label, attached with
liner) below the media sensor guide 3.
Align it so that the media sensor does not sense the
I-mark (black mark).
10
Close the print head.
11
Check the Sensor value.
If the value is 1.0 (V) higher than the "Low" level
value you have recorded, then this is the "High"
level value for the Gap sensor.
If the difference between the "High" and the "Low"
levels is less than 1.0, adjust the Emit and Receive
values so that the difference is more than 1.0, or
perform the adjustments again from step 2.
12
The standard values for the "High" and "Low" levels for the Gap sensor are as follows:
• Low (with only liner) ” 0.5 (V)
• High (media attached with liner) - Low • 1.0 (V)
13
If both "High" and "Low" levels comply with the standard value, press the right soft
button to confirm the value.
14
If you cannot adjust the sensor level, clean the sensor, check the connection of the
connector, and try again.
If the problem persists, replace the sensor. Refer to
details.
/
buttons to change
4 Checking and Performing Printer Adjustments
Section 5.4 Replacing the Media Sensor
CL4NX/CL6NX Service Manual
2
1
for
43