8 STORAGE

8.1 The dosimeters should be stored in a packing box in heated and
ventilated storehouses with air-conditioning at the ambient temperature
from + 5 to + 40 °C and relative humidity up to 80 % at + 25 °C
temperature, non-condensing. The storehouse should be free of acids, gas,
vapors of organic solvents, and alkali that may cause corrosion.
8.2 The location of the devices in storehouses should ensure their free
movement and access to them.
8.3 The dosimeters should be stored on the shelves.
8.4 The distance between the walls, the floor and the devices should not
be less than 100 mm.
8.5 The distance between the heating gadgets of the storehouse and the
devices should not be less than 0.5 m.
8.6 The average shelf life is not less than 6 years.
8.7 Additional information on storage, check during storage and
maintenance of the dosimeter is recorded in Appendices C, D, and H of this
OM.