3.1.3.4 Batteries switching off and their status control
The batteries should be switched off each time before the long-term
storage of the dosimeter. Do this as follows:
- switch the dosimeter off;
- open the lid of the battery compartment;
- remove the batteries;
- examine the battery compartment, check the contact clamps accuracy,
clean the battery compartment from contamination and contact clamps from
oxides;
- make sure there is no humidity, no salt spots on the surface of the
batteries, and no damages of the insulated coating.
3.2 Verification of the dosimeter
The dosimeters should be verified after manufacture, repair or during
use (periodically, at least once a year).
3.2.1 Verification operations
During verification, the operations presented in Table 3.2 should be
performed.