8 STORAGE

8.1 The radiometers should be stored in a packing box in heated and ventilated
storehouses with air-conditioning at the ambient temperature from + 5 to + 40 °C
and relative humidity up to 80 % at + 25 °C temperature, non-condensing. The
storehouse should be free of acids, gas, vapors of organic solvents, and alkali that
may cause corrosion.
8.2 The location of the radiometers in storehouses should ensure their free
movement and access to them.
8.3 The radiometers should be stored on the shelves.
8.4 The distance between the walls, the floor and the devices should not be less
than 100 mm.
8.5 The distance between the heating gadgets of the storehouse and the
radiometers should not be less than 0.5 m.
8.6 The average shelf life is not less than 6 years.
8.7 Additional information on storage, check during storage and maintenance
of the radiometer is recorded in Appendices B, C, and G of this OM.
116