Attaching the lamphouse and replacing the lamp

• To prevent electrical shock and damage to the microscope, always turn off the power switch
(flip it to the " " side) and unplug the power cord from the outlet before connecting or
disconnecting the lamphouse.
• To prevent burn injury, allow the lamp and the lamphouse to cool down sufficiently (for at least
30 minutes after the lamp is turned off), before replacing the lamp.
• Use the Nikon LV-LH50PC halogen lamphouse for the lamphouse.
• Use the Nikon LV-HL50W 12V 50W LONGLIFE halogen lamp or non-Nikon 12V 50W
SHORTLIFE halogen lamp (model OSRAM HLX 64610, OSRAM HLX 64611, or PHILIPS
7027) for the lamp. If you wish to buy these lamps, please contact your nearest Nikon
representative.
• Do not touch the glass surface of the lamp with bare hands. Fingerprints or grease on the bulb
surface will reduce the illumination intensity of the lamp. Wipe clean any fingerprints or grease
attached to the surface.
• Securely attach the lamphouse cover to the lamphouse after replacing the lamp. Never light
the lamp with the lamphouse cover removed.
• When you dispose of the replaced lamp, do not break it up. Instead, dispose of the used lamp
as special industrial waste or dispose of it according to the local regulations and rules.

4.4.1 Attaching the lamphouse

Before performing the following procedures, turn off the power switch of the microscope (flip
it to the " " side) and unplug the power cord from the wall outlet.
1.
Loosen the clamp screw on the upper side of the lamphouse connection port by using
the hexagonal screwdriver supplied with the microscope.
2.
Attach the episcopic lamphouse to the mount part on the upper-rear (or the diascopic
lamphouse to the mount part on the lower-rear) of the microscope, and push the
lamphouse into the limit.
Additionally, you can attach both the episcopic and diascopic lamphouses to the upper
and lower- mount part for the simultaneous illumination. Note that the lamp cable and
the TI-PS 100W Power Supply are required for the connection. (See 3.15 "L200ND
EPI/DIA simultaneous illumination".)
3.
Using the hexagonal screwdriver, tighten the clamp screw on the top of the connection
port of the lamphouse to secure the lamphouse.
4.
Plug the cable coming from the lamphouse into the lamp connector on the rear of the
microscope and tighten the ring of the connector to secure the connection.