4.5

Attaching the stage and the holder

The stage weighs about 10 kg.
Use caution not to drop it or get your hands and fingers caught.
1.
Lower the substage to the limit with the coarse focus knob.
2.
Place the stage on the substage and fix it by four M6 screws (supplied together with the
substage) using the 5 mm hexagonal wrench.
3.
Remove the fixing metals from the stage plate using 3 mm hexagonal wrench. The
fixing metals are attached to the stage with four hex socket head cap screws.
4.
Place the holder on the stage matching three claws on its bottom with the three claws
on the stage. Fix the holder with the clamp screw at the left side of the stage taking care
not to lift up the holder by tightening the clamp screw too much.
4.6

Attaching the eyepiece tube

Fully loosen the eyepiece tube clamp screw with the hexagonal screwdriver. Fit the
eyepiece tube on to the mount on the top of the microscope arm and tighten the eyepiece
tube clamp screw.

4.6.1 When removing the eyepiece tube

Hold the eyepiece tube when loosening the eyepiece tube clamp screw to prevent the
eyepiece tube from falling off.
4 Assembly
CAUTION
Screw holes for fixing the stage
Figure-45
38
IC Inspection Microscope ECLIPSE L200ND / L200N Instructions