• Choose a flat surface with little vibration.
• Choose a sturdy desk or table that is able to bear the weight of the instrument.
• The main body of this instrument weighs about 30 kg.
To install this instrument on a desk or a table, at least two people are required. Handle the
instrument carefully and do not drop it. Dropping may injure the person carrying it or
damage the instrument.
• Install this product in a place where it will not be a hazard in case of an unexpected event
such as an earthquake. In some locations, you may need to tie down this product to a
workbench or secure object so that it will not overturn or fall down.
Do not use this microscope in an environment where water can get into the microscope.
Do not install the microscope in a warm, humid location.
Install this microscope in a location where you can easily unplug the power cord from the
AC inlet of this microscope in emergency.
Provide enough space around the microscope referring to the layout diagrams on the next
page.
The installation conditions are as follows:
Area required : 710 mm (W) × 940 mm (D) or more
Power source : 100 to 240 V AC ±10% (1.2 A)
Humidity
: Relative humidity 85% at maximum (no condensation)
Operating environmental temperature : +5° to +35°C
Weight of the microscope
7 Cautions on moving the microscope
The microscope is a precision optical instrument. Handle it carefully and do not subject it to
a strong physical shock.
When moving the microscope, first remove the stage. Then, have two or more people
grasp the microscope from both sides.
(Information)
The main body of the L200ND/L200N weighs about 30 kg.
With the eyepiece tube, lamphouse and other parts (except stage)
attached, the microscope weighs about 35 kg.
Do not grasp the focusing knobs, eyepiece tube, lamphouse, stage mount, breath shielding
plate, etc., when carrying the microscope. They may come off and may cause serious
injury or malfunction.
Ask your nearest Nikon representative for the carrying rods of the microscope.
Before carrying the stage, attach the two fixing metals to hold the movement of the stage
plate. Use the smaller fixing metal to secure the side of the stage and then use the larger
fixing metal to secure the front of the stage.
Be careful not to pinch your fingers or hands when moving the microscope.
8 Cautions on assembling the microscope
Be careful not to pinch your fingers or hands during the assembly.
• The scratches or fingerprints on the lens surface will adversely affect the microscope's
image. Be careful not to scratch or touch the lens surfaces.
IC Inspection Microscope ECLIPSE L200ND / L200N Instructions