Error Handling

The following table lists warning and error messages and provides a description and possible cause.
Warning messages are returned when the camera cannot meet the full value of the request; error
messages are returned when the camera is unable to complete the request.
Table 10: Warning and Error Messages
Warning Messages
Camera Response
OK>
Warning 01: Outside of specification>
Warning 02: Clipped to min>
Warning 03: Clipped to max>
Warning 04: Related parameters adjusted>
Warning 07: Coefficient may be inaccurate
A/D clipping has occurred>
Warning 08: Greater than 1% of coefficients
have been clipped
Warning 09: Internal line rate inconsistent
with read out time>
Error Messages
Camera Response
Error 01: Internal error xx>
Error 02: Unrecognized command>
Error 03: Incorrect number of parameters>
Error 04: Incorrect parameter value>
Teledyne DALSA
Comment
Camera executed command
Parameter accepted was outside of specified operating range (e.g.
gain greater than ±10 dB of factory setting, or SSF below
specification).
Parameter was clipped to the current operating range. Use GCP or
GET to see value used.
Parameter was clipped to the current operating range. Use GCP or
GET to see value used.
Internal operating condition is adjusted to accommodate the entered
command. E.g. requesting exposure time longer than line time
automatically adjusts the line time to meet the exposure time
requirement.
In the region of interest (ROI) greater than 6.251% single or 1% of
averaged pixel values were zero or saturated.
Greater than 1% of FPN or PRNU coefficients have been calculated to
be greater than the maximum allowable and so were clipped.
Changing this parameter (e.g. vertical binning) has changed read out
time and that is greater than the internal SYNC
Comment
Where xx is a code list below.
Only output during power up.
Customer should contact Teledyne DALSA customer support.
Command is not valid.
Too many or too few parameters.
This response returned for:
Alpha received for numeric or visa versa
Float where integer expected
Not an element of the set of possible values. E.g., Baud Rate
Outside the range limit
55