Software Interface: How To Control The Camera - Dalsa Piranha HN HN-80-08k40-R User Manual

High sensitivity line scan ccd camera with enhanced qe in near-infrared
Table of Contents

Advertisement

18
4. Software Interface: How to
Control the Camera
All Piranha HN-xx camera features can be controlled through the serial interface. The camera can also be
used without the serial interface after it has been set up correctly. Functions available include:
Controlling basic camera functions such as gain and sync signal source.
Flat field correction.
Mirroring and readout control.
Generating a test pattern for debugging.
The serial interface uses a simple ASCII-based protocol and the PC does not require any custom software.
Note:
This command set has changes from previous Teledyne DALSA cameras. Do not assume that the
Piranha HN commands perform similarly to older cameras.
Serial Protocol Defaults
8 data bits
1 stop bit
No parity
No flow control
115,200 kbps (fixed)
Camera does not echo characters
Command Format
When entering commands, remember that:
A carriage return <CR> ends each command.
A space or multiple space characters separate parameters. Tabs or commas are invalid parameter
separators.
Upper and lowercase characters are accepted
The backspace key is supported
The camera will answer each command with either <CR><LF> "OK >" or <CR><LF>"Error xx: Error
Message >" or <CR><LF>"Warning xx: Warning Message >". The ">" is used exclusively as the last
character sent by the camera.
The following parameter conventions are used in the manual:
= integer value
i
= real number
f
= member of a set
m
= string
s
= tap id
t
= pixel column number
x
03-032-20135-00
Piranha HN RoHS User Manual
Teledyne DALSA

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Piranha hn hn-80-08k40

Table of Contents