Introduction

Updating UTX878
Instrument Software
Calibration
In the CALIB menu, you can calibrate and trim the analog outputs and
inputs and check other meter functions. This chapter also covers
updating UTX878 software over the RS232 interface.
Before performing calibration of the UTX878, be sure the following
equipment is available:
0-30 VDC, 100 mA variable power supply
Current meter capable of precisely measuring 4 to 20 mA current
levels
PC with RS232 Cable and Hyperterminal software (available on
Windows operating systems)
While following the programming instructions, refer to Figure A-3 on
page A-3 of Appendix A, Menu Maps.
1. To set up the RS232, connect one end of a 9-pin RS232 cable to
the COM-1 serial port on a PC and the other end to terminal block
TB4 on the UTX878, as described in Chapter 1, Installation.
2. On your PC, set up the Hyperterminal program.
Note: The use of Hyperterminal is shown here as an example. If you
are using a different communications software, see its manual
for detailed instructions.
a. From the PC Start Menu, click Programs/Accessories/
Communications/Hyperterminal to open the Hyperterminal
window.
b. If the call is not connected, click on New Connection and enter
a name. Click OK.
c. In the Connect To window, select COM 1 as the desired port.
d. In the Properties window, set the following parameters:
Bits per second: 19200
Data bits: 8
Parity: None.
Stop Bits: 1
Flow control: None
March 2004
5-1