March 2004
Entering Reynolds
Correction Data (cont.)
2-14
3. The menu varies, depending on your selection in Step 2.
•
If you select Off, no further choices are available.
•
If you select Single, the UTX878 will select and automatically
display the Kinematic Viscosity. To change the value, press [ENT].
Use the arrow keys to change the value (available in document
#914-004, Sound Speeds and Pipe Size Data), and press [ENT].
•
If you select Table, the screen displays three options: Units, Rows
Used and Edit Table.
a. If you scroll to Units and press [ENT], the screen displays three
more options: Velocity, Soundspeed and Diagnostic. If you
select either Velocity or Soundspeed and press [ENT], the
screen displays the measurement units (either metric or
English). Press [ESC] or [ENT] to return to the previous
screen. But if you scroll to Diagnostic and press [ENT], the
meter asks for the type of signal to be used, Signal Strength Up
or Signal Strength Dn. Scroll to the appropriate signal, and
press [ENT].
b. If you select Rows Used, the program asks for the number of
rows you wish to use. Enter the desired number (from 2 to 20)
and press [ENT].
c. If you select Edit Table, the table opens with a series of rows.
Scroll to the desired row, and press [ENT].
d. For each row, the screen displays the Reynolds Correction
number (X) and the Kinematic Viscosity (KV). If you wish to
change either value, scroll to the value and press [ENT]. Use
the arrow keys to change the value (available in document
#914-004, Sound Speeds and Pipe Size Data), and press
[ENT].
e. Repeat steps c and d until you have programmed all of your
available data (from 2 to 20 rows).
Press [ESC] until you return to the PROG menu, or continue pressing
[ESC] until the display screen reappears.
Programming Site Data