Entering Reynolds
Correction Data (cont.)
2-14
3. The menu varies, depending on your selection in Step 2.
If you select Off, no further choices are available.
If you select Single, the UTX878 will select and automatically
display the Kinematic Viscosity. To change the value, press [ENT].
Use the arrow keys to change the value (available in document
#914-004, Sound Speeds and Pipe Size Data), and press [ENT].
If you select Table, the screen displays three options: Units, Rows
Used and Edit Table.
a. If you scroll to Units and press [ENT], the screen displays three
more options: Velocity, Soundspeed and Diagnostic. If you
select either Velocity or Soundspeed and press [ENT], the
screen displays the measurement units (either metric or
English). Press [ESC] or [ENT] to return to the previous
screen. But if you scroll to Diagnostic and press [ENT], the
meter asks for the type of signal to be used, Signal Strength Up
or Signal Strength Dn. Scroll to the appropriate signal, and
press [ENT].
b. If you select Rows Used, the program asks for the number of
rows you wish to use. Enter the desired number (from 2 to 20)
and press [ENT].
c. If you select Edit Table, the table opens with a series of rows.
Scroll to the desired row, and press [ENT].
d. For each row, the screen displays the Reynolds Correction
number (X) and the Kinematic Viscosity (KV). If you wish to
change either value, scroll to the value and press [ENT]. Use
the arrow keys to change the value (available in document
#914-004, Sound Speeds and Pipe Size Data), and press
[ENT].
e. Repeat steps c and d until you have programmed all of your
available data (from 2 to 20 rows).
Press [ESC] until you return to the PROG menu, or continue pressing
[ESC] until the display screen reappears.