FRTD commands

:TYPE <name>
[:SENSe[1]]:TEMPerature:FRTD:TYPE <name>
Parameters
Query
Description
:RZERo <NRf>
[:SENSe[1]]:TEMPerature:FRTD:RZERo <NRf>
Parameters
Query
Description
<name> =
PT100
D100
F100
PT3916
PT385
USER
:TYPE?
This command is used to select the FRTD standard and other related factors.
When one of the parameters other than USER is selected, the multimeter
defaults to the following factors:
Type
Alpha
PT100
0.003850
D100
0.003920
F100
0.003900
PT385
0.003850
PT3916
0.003916
Changing alpha (see :ALPHa), beta (see :BETA), delta (see :DELTa), or Ω at
0°C (see :RZERo) automatically changes the type to USER (see :TYPE).
<NRf> = 0 to 10,000
:RZERo?
This command is used to check and/or change the resistance at 0°C.
Remember that if you change the present resistance value, the FRTD TYPE
will change to USER.
Specify FRTD parameters
Selects default parameters for the PT100
standard (ITS-90)
Selects default parameters for the D100
standard (ITS-90)
Selects default parameters for the F100 standard
(ITS-90)
Selects default parameters for the PT3916
standard (IPTS-68)
Selects default parameters for the PT385
standard (IPTS-68)
Selects user-defined standards
Query type of FRTD sensor
Beta
Delta
0.11100
1.50700
0.10630
1.49710
0.11000
1.49589
0.11100
1.50700
0.11600
1.50594
Specify resistance at 0°C
Specify FRTD resistance at 0°C
Query value of RZERo
Ω at 0°C
100Ω
100Ω
100Ω
100Ω
100Ω