Trigger Model (Gpib Operation) - Keithley 2010 User Manual

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

4-24
Remote Operation

Trigger model (GPIB operation)

The following paragraphs describe how the Model 2010 Multimeter operates over the GPIB
bus. The flowchart in
trigger model because operation is controlled by SCPI commands from the Trigger subsystem
(see Section 5 for more information). Key SCPI commands are included in the trigger model.
Figure 4-11
Trigger model (remote
operation)
Figure 4-11
summarizes operation over the bus. The flowchart is called the
:ABOrt
*RCL
:SYST:PRES
Language Change
:INIT (:IMM)
Idle
or
and
:INIT:CONT ON
Initiate
Yes
Event
Control
Detection
Source
:Trigger:Source Immediate
:Trigger:Source External
:Trigger:Source Timer
:Trigger:Source Manual
:Trigger:Source BUS
:Trigger:Delay <n>
Delay
:Trigger:Delay:AUTO <b>
No
?
:Trigger:Signal
Device
Action
(see Figure 4-12)
No
:INIT (:IMM)
Yes
or
:INIT:CONT ON
?
No
Yes
Another
Trigger
?
:Trigger:Count <n> Infinite
Output
Trigger
No
Yes
Another
Sample
?
:Sample:Count <n>