Keithley 2010 User Manual page 149

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

5-12
SCPI Command Reference
Table 5-6 (cont.)
SENSe command summary
Command
:VOLTage:AC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:STATe <b>
:STATe?
:DETector
:BANDwidth <NRf>
:BANDwidth?
Description
Path to configure AC voltage.
Set integration rate (line cycles; 0.01 to 10).
Query line cycle integration rate.
Path to configure measurement range:
Select range (0 to 757.5).
Query range.
Enable or disable auto range.
Query auto range.
Specify reference (-757.5 to 757.5).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolution (4 to 7).
Query resolution.
Path to configure and control the filter.
Select filter type: (MOVing or REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Path to configure bandwidth:
Specify bandwidth (3 to 300e3).
Query bandwidth.
Default
SCPI
parameter
1
757.5
ON
0
OFF
6
(Note)
10
OFF
30

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents