:RANGe commands

[:UPPer] <n>
[:SENSe[1]]:CURRent:AC:RANGe[:UPPer] <n>
[:SENSe[1]]:CURRent[:DC]:RANGe[:UPPer] <n>
[:SENSe[1]]:VOLTage:AC:RANGe[:UPPer] <n>
[:SENSe[1]]:VOLTage[:DC]:RANGe[:UPPer] <n>
[:SENSe[1]]:RESistance:RANGe[:UPPer] <n>
[:SENSe[1]]:FRESistance:RANGe[:UPPer] <n>
[:SENSe[1]]:VOLT[:DC]:STERminals:RANGe[:UPPer] <n> Set measurement range for sense terminals
Parameters
Query
Description
<n> =
0 to 3.1
0 to 757.5
0 to 1010
0 to 10.1
0 to 120e6
DEFault
MINimum
MAXimum
:RANGe[:UPPer]?
:RANGe[:UPPer]? DEFault
:RANGe[:UPPer]? MINimum
:RANGe[:UPPer]? MAXimum
This command is used to manually select the measurement range for the
specified measurement function. The range is selected by specifying the
expected reading as an absolute value. The Model 2010 will then go to the
most sensitive range that will accommodate that expected reading. For exam-
ple, if you expect a reading of approximately 50mV, simply let the parameter
(<n>) = 0.05 (or 50e-3) in order to select the 100mV range.
SCPI Command Reference
Set measurement range for ACI
Set measurement range for DCI
Set measurement range for ACV
Set measurement range for DCV
Set measurement range for Ω2
Set measurement range for Ω4
Expected reading is amps (ACI and DCI)
Expected reading is AC volts (ACV)
Expected reading in DC volts (DCV)
Expected reading in DC volts (DCV) for sense
terminals
Expected reading is ohms (Ω2 and Ω4)
3.03 (ACI and DCI)
757.5 (ACV)
1010 (DCV)
101e6 (Ω2 and Ω4)
0 (All functions)
Same as DEFault
Query measurement range
Query *RST default range
Query lowest measurement range
Query highest measurement range