5-18
SCPI Command Reference
Table 5-8
SYSTem command summary
Command
:SYSTem
:PRESet
:POSetup <name>
:POSetup?
:FRSWitch?
:VERSion?
:ERRor?
:AZERo
:STATe <b>
:STATe?
:KEY <NRf>
:KEY?
:CLEar
:BEEPer
:STATe <b>
:STATe?
:LOCal
:REMote
:RWLock
:KCLick <b>
:KCLick?
:LFRequency?
Note: Clearing the Error Queue: Power-up and *CLS – Clears the Error Queue.
Table 5-9
TRACe command summary
Command
:TRACe:DATA
:CLEar
:FREE?
:POINts <NRf>
:POINts?
:FEED <name>
:CONTrol <name>
:CONTrol?
:FEED?
:DATA?
*:SYSTem:PRESet and *RST have no effect on the commands in this subsystem.
Description
Return to :SYST:PRES defaults.
Select power-on setup: (RST, PRESet or SAV0).
Query power-on setup.
Query INPUTS switch (0=rear, 1=front).
Query rev level of SCPI standard.
Query (read) Error Queue.
Path to set up autozero.
Enable or disable autozero.
Query autozero.
Simulate key-press (1 to 31; see
Query the last "pressed" key.
Clears messages from the Error Queue.
Path to control beeper.
Enable or disable beeper.
Query state of beeper.
Take 2010 out of remote and restore operation of front panel
controls (RS-232 only).
Place 2010 in remote (RS-232 only).
Lockout front panel controls (RS-232 only).
Turn the keyclick on/off.
Query the keyclick status.
Query power line frequency.
*RST, :SYSTem:PRESet, and :STATus:PRESet – No effect.
Description
Use :TRACe or :DATA as root command.
Clear readings from buffer.
Query bytes available and bytes in use.
Specify size of buffer (2 to 1024).
Query buffer size.
Select source of readings (SENSe[1], CALCulate[1], NONE).
Select buffer control mode (NEVer or NEXT)
Query buffer control mode.
Query source of readings for buffer.
Read all readings in the buffer.
Figure
5-10).
Default
SCPI
parameter
(Note)
ON
ON
ON
Default
SCPI
parameter*

Hide quick links: