[Sense[1]] Subsystem; Function Command - Keithley 2010 User Manual

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

[SENSe[1]] subsystem

:FUNCtion Command

:FUNCtion <name>
[:SENSe[1]]:FUNCtion <name>
Parameters
Query
Description
The Sense 1 Subsystem is used to configure and control the measurement
functions of the Model 2010. A function does not have to be selected before
you program its various configurations. A function can be selected any time
after it has been programmed. Whenever a programmed function is selected,
it assumes the programmed states.
<name> = 'CURRent:AC'
'CURRent[:DC]'
'VOLTage:AC'
'VOLTage[:DC]'
'RESistance'
'FRESistance'
'PERiod'
'FREQuency'
'TEMPerature'
'DIODe'
'CONTinuity'
:FUNCtion?
The :FUNCtion command is used to select the measurement function of the
instrument. Note that parameter names are enclosed in single quotes (').
However, double quotes (") can instead be used. For example:
:func 'volt' = :func "volt"
Each measurement function remembers its own unique setup configuration,
such as range, speed, resolution, filter, and rel. This eliminates the need to re-
program setup conditions each time you switch from one function to another.
SCPI Command Reference
Select measurement function.
'Select AC Current
'Select DC Current
'Select AC Voltage
'Select DC Voltage
'Select two-wire Resistance
'Select four-wire Resistance
'Select Period
'Select Frequency
'Select Temperature
'Select Diode Testing
'Select Continuity Testing
Query presently programmed function.
5-37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents