The Model 2010 uses the 1kΩ range to measure circuit continuity. After selecting continuity,
the unit prompts you for a threshold resistance level (1Ω-1000Ω). The Model 2010 alerts you
with a beep when a reading is below the set level.
To measure the continuity of a circuit, press SHIFT then CONT, set the threshold resistance
level, and connect the circuit.
NOTE

Connections

Connect the circuit you want to test to the INPUT HI and INPUT LO terminals of the
Model 2010. The test current flows from the INPUT HI as shown in
Figure 2-10
Continuity measurements

Threshold resistance level

You can define a threshold resistance from 1Ω to 1000Ω. The factory setting is 10Ω. Follow
these steps to define the resistance level:
1.
2.
3.
Continuity has a non-selectable reading rate of FAST (0.1 PLC).
Model 2010
REM
STEP
SCAN
CH1
CH2
CH3
CH4 CH5
CH6
CH7
CH
CH9
CH1
MATH
TALK
8
0
REAR
LSTN
SRQ
SHIFT
4W
TIMER
HOLD TRIG
FAST
MED
SLOW
REL
FILT
AUTO
ERR
BUFFER
STAT
2001 MULTIMETER
Note: Source current flows from the INPUT
HI to INPUT LO terminals.
Press SHIFT then CONT.
keys to choose a numerical place and use the L and M keys to
Use the
and
increment or decrement the digits. Enter a value from 1 to 1000.
Press ENTER to confirm your setting.
Figure
2-10.
Resistance
Under Test