Display Subsystem - Keithley 2010 User Manual

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

5-26
SCPI Command Reference
:IMMediate
:CALCulate3:IMMediate
Description
Program

DISPlay subsystem

:ENABle <b>
:DISPlay:ENABle <b>
Parameters
Query
Description
When the configuration of the limit test is changed, the next reading is
evaluated according to the new test configuration. If the instrument is not in
a continuous measurement mode (e.g., waiting for a manual trigger), the test
will not be performed until the next reading conversion occurs.
This action command lets you re-process the current input data to test new
limits. For example, assume the instrument is in a non-continuous
measurement mode and requires a manual trigger to cause the next reading
conversion. Changing the test limits will not affect the last test result.
However, sending the :IMMediate command reprocesses the data and
evaluates the reading according to the new test limits. Note that sending the
:IMMediate command does not initiate a reading conversion.
PRINT #1, "output 16;:trig:sour bus"
SLEEP 3
PRINT #1, "output 16;:calc3:imm"
The commands in this subsystem are used to control the display of the
Model 2010 and are summarized in
<b> =
0 or OFF
1 or ON
:ENABle?
This command is used to enable and disable the front panel display circuitry.
When disabled, the instrument operates at a higher speed. While disabled, the
display is frozen.
All front panel controls (except LOCAL) are disabled. Normal display
operation can be resumed by using the :ENABle command to enable the
display or by putting the Model 2010 into local mode (press LOCAL).
Perform CALC3
' Place 2010 in one-shot mode
' Wait three seconds
' Re-perform limit test
Table
5-3.
Control display circuitry
Disable display circuitry
Enable display circuitry
Query state of display