Feed Command; Data? Command - Keithley 2010 User Manual

Hide thumbs Also See for 2010:
Table of Contents

Advertisement

5-72
SCPI Command Reference

:FEED command

:FEED <name>
:TRACe:FEED <name>
Parameters
Query
Description
:CONTrol <name>
:TRACe:FEED:CONTrol <name>
Parameters
Query
Description

:DATA? command

:DATA?
:TRACe:DATA?
Description
<name> = SENSe[1]
CALCulate[1] Put calculated readings in buffer
NONE
:FEED?
Query buffer feed
This command is used to select the source of readings to be placed in the
buffer. With SENSe[1] selected, raw readings are placed in the buffer when
storage is performed. With CALCulate[1] selected, calculated math readings
(MX+B or PERCent or NONE) are placed in the buffer.
With NONE selected, no readings are placed in the buffer when storage is
performed over the bus.
<name> = NEVer
NEXT
:CONTrol?
Query buffer control
This command is used to select the buffer control. With NEVer selected,
storage into the buffer is disabled. With either of the other selections, storage
is performed as long as buffer feed is not set for NONE (see :TRACe:FEED
NONE). When NEXT is selected, the storage process starts, fills the buffer,
and then stops. The buffer size is specified by the :POINts command.
When this command is sent and the Model 2010 is addressed to talk, all the
readings stored in the buffer are sent to the computer. The format that
readings are sent over the bus is controlled by the :FORMat subsystem.
Specify readings source
Put raw readings in buffer
Put no readings in buffer
Specify buffer control
Disables buffer storage
Fills buffer and stops
Send buffer readings

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents